Tell me more about the demodulated echo signals
In pulsed ultrasound Doppler
velocimetry the sampling volume contains not a unique particle but a lot of
small particles having most of the time different shapes, different sizes and
different acoustic impedances. As all of these particles contribute to generate
a unique value of the amplitude of the echo for each emission, the evolution
of the echo amplitude will fluctuate. The reason comes from the fact that some
particles enter in the sampling volume and others leave it. This fluctuation
in amplitude will remain after the demodulation process and will be present
on the demodulated Doppler signal, often called I and Q.
All the Doppler frequency shifts induced by the movement of the particles are
combined together. If all the particles do not have the same velocity (in amplitude
and in direction) the demodulated signals will contain many Doppler frequencies.
The resulting demodulated signals I and Q contain therefore the combination
of these two phenomena. In order to have an idea of the aspect of a real demodulated
Doppler signal corresponding to one gate, the figure below illustrates its evolution.

The best way
to analyze the frequency content of the demodulated echo signal issued from
one gate is to compute its power spectrum. The power spectrum gives information
on the distribution of the measured Doppler frequencies and their relative influences
on the computed mean Doppler frequency, which is computed when the velocimeter
displays the velocity profile.
The DOP1000 could compute the power spectrum of a data series by means of an
FFT algorithm. The data series is formed of samples taken on the demodulated
I and Q signals for a selected depth which corresponds to one gate. Before the
computation of the power spectrum, the data series is filtered by a high pass
filter which removes all the stationary components contained in the demodulated
signals. In order to make the reading of the frequency scale more easy the frequency
scale is converted in velocity by using the standard Doppler formula. The ordinate
gives the relative amplitude, in a logarithmic scale, of the power spectrum.
Why using the power spectrum of one gate
When once looks at the velocity profile, all the gates give a single value of velocity. These values are the result of a computation of the mean Doppler frequency, which is the real mean value (non biased) of the power spectrum. This means that no information is given about the distribution of the Doppler energy. The same value of velocity could result from many different frequency distributions. The display of the complete power spectrum is a good method to increase the knowledge on the measured velocity values.
How to use the information contained in the power spectrum
The following examples will display different situations where the power spectrum could increase the knowledge in the measured velocity values.

Figure 1
The first example in figure 1 shows the computed power spectrum from a gate placed in the middle of a tube where a liquid is flowing. As shown in the power spectrum the sampling volume does not contain a single Doppler frequency. The width of the peak is related to the number of Doppler frequencies present in the sampling volume.

Figure 2
In the second example, in figure 2, the sampling volume has been moved closer to the wall of the tube. The width of the Doppler peak is now much larger which means that much more different velocities are present in the sampling volume. The power spectrum reveals also a small influence of the movements of the walls. As the power spectrum is computed from the high-passed filtered data values the amplitude of the power spectrum at the origin is always zero.

Figure 3
The third example,
in figure 3, considers a rotating cylinder filled of liquid. The ultrasonic
beam crosses the cylinder perpendicularly to the axis of the cylinder. The sampling
volume is placed inside the cylinder in a position near the wall.
The power spectrum reveals a very strong influence of the movements of the walls
of the cylinder despite the sampling volume does not touch a wall. This situation
may appear if some of the multiple ultrasonic reflections may coincide with
the sampling time of the echo. In such a case the velocity value is corrupted
by the Doppler effect induced by the movement of the walls. The display of the
power spectrum indicates clearly that the mean Doppler frequency computed is
in fact the result of the mean value of two different velocity components, one
coming from the movements of the walls and the other coming from the movements
of the particles contained in the liquid.