Measurement of the Insertion Loss
The insertion loss characterize the sensitivity of a transducer, and is given by :

In order to measure the
insertion loss, an ultrasonic burst is emitted by a transducer. The same transducer
receives an echo issued from a plane perpendicular target, which acoustic impedance
differs very much from the liquid used as a transmitting media.
The voltage ratio between the echo and the emitted burst, expressed in logarithm,
is called Insertion Loss (IL).
Signal Processing's transducers have an IL between 10 and 20 dB, depending on the type of transducer.
To measure the Insertion Loss of your transducer, you need the following equipment :
- 1 burst generator
- 1 oscilloscope
- 2 coaxial cable (50 W) + 1 T
- 1 water container, with a transducer holder
- 1 target (stainless steel preferred)
Set up
Burst generator :
Fburst = Ftransducer
Amp = 10 Vpp / 50 W
Nb burst = 4
Space between burst =
10000 Cycles
Connect the transducer, the
burst generator and the oscilloscope as shown in the figure below.
You can trig the Echo signal by using the TTL Output signal coming from the
burst generator (if available).

Warning
To prevent any damage to
the burst generator, we recommend to check the specifications of the instrument
in order to verify that it can support the connection of a very low impedance
(<50 ohm) during a long time.
Measuring procedure
1. Be sure that the contact
between the front surface of the transducer and the water is good (no air bubbles).
2. Be sure that the transducer is perpendicular to the target (max echo voltage).
3. Measure the voltage
of the emitted burst.
4. Measure the voltage of the echo.
5. Compute the Insertion Loss.
Important notices
The distance between the transducer and the target must be enough in order to avoid the disturbances generated by the ringings, but not too long to be sure to collect all the emitted energy (divergence of the ultrasonic beam).
The thickness of the
target must also be considered.
To avoid any interference
generated by second echo, coming from the far wall of the target,
a minimum thickness is required.
As the amount of the
energy that penetrates inside the target depends on the difference in acoustic
impedance between the target and the transmitting media. It is recommended to
use a target with an acoustic impedance as high as possible.
The following table gives recommended values for the thickness of the target (Tk) and the distance between the transducer and the target (D) for different transducers.
|
F
[MHz]
|
D [mm] |
Tk [mm] |
|
0.5
|
60
|
20
|
|
1
|
40
|
10
|
|
2
|
40
|
10
|
|
4
|
30
|
10
|
|
8
|
30
|
5
|
|
10
|
30
|
5
|
For D : 4 cycle burst, time between the emitted burst and the received echo of 40 ms, and a sound velocity of 1500 m/s (water).